Accuracy versus Resolution in Analyzing System Errors
Dataforth Corporation
Monday, December 01 2008
advertisement:
Distinguishing between accuracy and resolution can
be misinterpreted in determining system needs.
To ensure a system’s accuracy meets required needs, system error
budgets must be an integral part of system design. Considerations
should include necessary levels of accuracy for system elements, as
well as such issues as compatibility between software algorithm calculations
and measurement accuracy – meaning resolution must
also be taken into account.
Figure 1: Typical Instrumentation Signal Flow. Sensor signals are conditioned with signal conditioning modules, selected, and converted into a usable number either for analytical process control or observation.
Accuracy is the degree of absolute correctness of a measurement
device; resolution is the smallest number that the device can
display or record. In the following examples, the digital device
quantizing error (±1 bit minimum) in the least significant digit is
assumed to be zero. Remember that a measurement device with a
specified accuracy of ±0.015% actually gives an output between
0.99985 and 1.00015 times the actual true value.
Figure 2: A Conceptual System used to convert an analog signal to digital representation.
1. Measure a voltage source known to be exactly 5.6430 volts
with a digital voltmeter that is (somehow) 100% accurate but has
only 3 display digits, defined as “3-digit resolution.” The reading is
5.64 volts, which does not represent the actual voltage value
although both the source and the instrument are 100% accurate.
Resolution here is 10mV.
Table 1. Accuracy Calculations for Figure 1.
2. Measure the precise 5.643-volt source using a 5-digit display
digital voltmeter with a specified accuracy of ±0.015%. The reading
is between 5.6421 and 5.6438. This is closer to the actual voltage
(5.6430), but still not 100% accurate. Resolution in this case is 1 mV.
Measuring 1 volt within ±0.015% accuracy requires a 6-digit
instrument capable of displaying five decimal places. The fifth decimal
place represents a resolution of 10 microvolts. Using any instrument with less than 6 digits, “accuracy”
is determined by the resolution of the
reading instrument and the acceptance of
the observer.
Table 1 displays some different system
“accuracy” correction calculations. Since
errors are random and have ± values, RMS
calculations are often used as opposed to
worst-case maximum and minimum. RMS
error is defined as the square root of the
sum of each error squared, ˆ {(E1)2
+(E2)2 + (E3)2}.
Analog-to-digital converters (ADC) are
advertised as having “n” bit resolution –
often misunderstood to mean accuracy.
The effective accuracy of an n-bit ADC is
not equal to ADC resolution, which is
defined as approximately 1/(2n-1). Figure
2 shows a conceptual system used to convert
an analog signal to digital representation.
Semiconductor switches select analog
input signals, which are captured and held
in a sample and hold amplifier function
block (SHA). An n-bit counter then begins
to count, and the counter contents are converted
to an analog voltage using switched
resistors or current sources. When the analog
and SHA signals are equal, counting
stops and the counter contents become
available as a digital representation of the
sampled analog input value. The process,
however, includes sources of error that collectively
degrade true accuracy.
Errors associated with typical ADC
schemes:
Sampling Speed. From Nyquist Sampling
Theory, if the analog signal changes rapidly,
the ADC must sample at least twice
as fast as the changing input. Sampling
slower than one-half the signal frequency
will result in inaccurate readings.
Input Multiplexer. Input multiplexer circuits
may have OpAmp buffers on each
input line that could introduce errors in
voltage offset, current bias, and linearity.
In addition, multiplexers can create
cross talk between channels and signal
attenuation.
Sample and Hold Amplifier. This function
is an OpAmp-based circuit with components
designed to switch, buffer, and
hold the sampled analog voltage value.
Consequently linearity, gain, power supply
shifts, voltage offsets, charge injection,
and input bias currents will contribute
errors.
Converter. In the counter, comparator,
and ADC circuit there are such errors as
overall linearity, quantizing error (uncertainty
in the least significant bit), and
power supply shifts.
Temperature. All analog circuit functions
within the ADC unit are subject to temperature
errors.
Obviously, many factors, including resolution,
must be considered to determine
overall system accuracy. Often, the errors in
a total system error budget are predominantly
from industrial sensors used in
process control and data acquisition systems
because they can have accuracies
much lower than SCMs or ADC units.
This article was written by John Lehman,
Engineering Manager, at Dataforth Corporation.
For more information, visit http://info.hotims.com/15144-121.
Subscribe today to receive the INSIDER, a FREE e-mail newsletter from NASA Tech Briefs featuring exclusive previews of upcoming articles, late breaking NASA and industry news, hot products and design ideas, links to online resources, and much more.