The figure depicts a replaceable microfluidic cartridge that is a component of a miniature biosensor that detects target deoxyribonucleic acid (DNA) sequences. The biosensor utilizes (1) polymerase chain reactions (PCRs) to multiply the amount of DNA to be detected, (2) fluorogenic polynucleotide probe chemicals for labeling the target DNA sequences, and (3) a high-sensitivity epifluorescence-detection optoelectronic subsystem.
The design and fabrication of the replaceable microfluidic cartridges meets these challenges. The cartridges are made from a combination of materials, based on micromachined silicon substrates. The thermal-expansion-mismatch problem was solved by use of thick (3 mm) silicon substrates and anodically bonded thick (0.7 mm) covers made of borosilicate float glass. The fluorescence signal is enhanced by use of multipass optics, an essential component of which is a reflective film of chemical-vapor-deposited aluminum on a square glass plate affixed to the bottom of the cartridge by use of epoxy. The interior surfaces of the fluidic channels are coated with dodecyltriethoxysilane.
The design of the cartridge has been optimized along with that of the rest of the biosensor for detection of target DNA in a microgravitational or normal gravitational setting. In a test, the biosensor was found to be capable of reliably detecting 600,000 copies of the human ß-actin gene after as few as five PCR cycles.
This work was done by Kevin Francis and Ron Sullivan of Systems and Processes Engineering Corp. for Johnson Space Center. For further information, contact:
SPEC
101 West 6th Street, Suite 200
Austin, TX 78701-2932
Phone: (512) 479-7732
Fax: (512) 494-0756
E-mail:
This e-mail address is being protected from spambots. You need JavaScript enabled to view it
Refer to MSC-23348.